Manitoba Dental X-Ray Rules: What Changes on January 31, 2027

The short answer

In Manitoba, dental X-ray equipment falls today under the X-Ray Safety Regulation (Man. Reg. 341/88 R), administered by CancerCare Manitoba's Radiation Protection Services. Owners must register each unit and get location approval; the regulation has no quality assurance rule. From January 31, 2027, The Radiation Protection Act requires owners of ionizing radiation equipment intended for a health care purpose to implement a QA program. No regulations under the Act are published yet, including the energy range that sets which equipment it covers and any QA program standards.

Who regulates what

Profession regulators

X-ray equipment authority

Radiation Protection Services, CancerCare Manitoba (CCMB), a department of Medical Physics, says it is responsible to the Government of Manitoba and administers the regulation under The Public Health Act.

Federal layer: Health Canada says federal law covers the sale, lease and import of X-ray equipment, while provinces may set rules for its installation and use.

What the law requires

Today: the X-Ray Safety Regulation

  • Registration: within 30 days of becoming the owner, with notice within 30 days of ceasing to own a unit (section 12).
  • Location approval: the minister's prior approval of a permanent X-ray location, with scale drawings, and again for a different machine type, more machines or substantial alterations (section 13).
  • Dosimeters: worn by X-ray workers while on duty (section 11).
  • Inspections: allow inspector entry, questions and technical tests (section 16).

The regulation has no quality assurance provision and no inspection frequency. CCMB adds that a shielding specification is needed before construction, renovation or relocation.

From January 31, 2027: The Radiation Protection Act

The Radiation Protection Act, S.M. 2015, c. 41 repeals the regulation (section 61). It covers equipment that emits ionizing radiation within an energy range to be set by regulation (section 3(1)). Owners of covered equipment must:

  • operate it only if it is registered or exempt and, unless it is portable (it can be carried by hand to where it is operated and is not mobile equipment), only in a location the director has approved (sections 2(1), 5(1), 9(1) and 9(3));
  • apply to the director to amend an approval before changes that may affect patient dose or image quality, and before replacing or adding equipment (section 13(1));
  • maintain it so it properly performs its intended function and is unlikely to cause an overexposure (section 18(1));
  • implement a QA program for equipment intended for a health care purpose, such as diagnosis, imaging or therapy, following any program standards in the regulations (section 19);
  • provide dosimeters and keep dose and exposure records in accordance with the regulations (sections 25, 27 and 28).

Existing registrations and permanent location approvals carry over (sections 57 and 58). Owners who apply within 30 days after the Act takes effect are not caught by the registration and location bans until a decision is made (section 59).

Still unknown. As of November 1, 2026, Manitoba's annual regulations index listed no regulation under the Act, so the energy range, QA standards, any test frequency and any record period are unpublished. The Act does not name its administering body (sections 4 and 30).

Key dates

  • August 29, 1988: the X-Ray Safety Regulation is registered. Manitoba's laws site still lists it as the first version.
  • November 5, 2015: The Radiation Protection Act receives assent.
  • March 11, 2026: a proclamation under Order in Council 35/2026 fixes the in-force date.
  • January 31, 2027: the Act takes effect and the regulation is repealed.
  • No date yet: regulations under the Act.

A practical checklist for your clinic

  1. Confirm registrations. Each intraoral and panoramic unit should be registered with CCMB (section 12(1); carried over by section 57) and carry the registration label CCMB issues.
  2. Check approvals after changes. CCMB says registration does not approve a location. New locations and added or different machines need prior approval (section 13).
  3. Act on CCMB inspection findings. CCMB's overview says deficiencies must be addressed within 30 days of receiving its report, and owners must comply with all recommendations to keep ministerial permission to operate.
  4. Write down the QC and service work you already do: devices, checks, frequency, who does them and where records are kept. Good practice ahead of sections 18 and 19.
  5. Know how you would produce electronic records. From January 31, 2027, inspectors may require electronic records as a printout, in an electronically readable format, or both (section 31(3)).
  6. Watch for the regulations. Good practice: check the annual regulations index.

Where Phantom XY and Phantom Cloud fit

Phantom XY System (test hardware) and Phantom Cloud (web software) run an automated digital QA test of about two minutes on intraoral and panoramic X-ray devices, score each result with a Q-Score, and keep the results in one place. Results, images and logbook entries export to PDF. Phantom Cloud sends missed-test and failure alerts and includes a Digital X-Ray Logbook. Manitoba has not yet published its QA program standards, so MVD makes no claim that Phantom meets them. The owner remains responsible for the program and for meeting the standards once they are published.

Common questions

Does Manitoba require daily dental X-ray QA?

Not under the X-Ray Safety Regulation, which has no quality assurance provision. From January 31, 2027, The Radiation Protection Act requires a QA program that follows any standards set in the regulations. None were published as of November 1, 2026, so no test frequency is set under either law. Dentists should also check with the Manitoba Dental Association.

What will a Manitoba QA program have to include?

Section 19(2) says the program is meant to ensure results of satisfactory quality when equipment is used for its intended purpose. Section 19(3) ties it to standards in the regulations, which may adopt an existing code or standard by reference (section 55(3)). Until regulations are published, the content is unknown.

Who inspects dental X-ray equipment in Manitoba, and how often?

Radiation Protection Officers at CancerCare Manitoba, whom CCMB describes as provincially appointed inspectors. CCMB lists mandatory compliance inspections on a guidance schedule of every 3 to 5 years. That interval is CCMB guidance, not a number set in the regulation. From January 31, 2027, the minister may appoint inspectors under the Act (section 30(1)), which does not name their employer.

Do registered X-ray units need to be registered again in 2027?

Not if they are registered under the current regulation. Section 57 deems equipment registered under the old regulation on the day before the Act takes effect to be registered under the Act. Section 58 continues permanent location approvals. An unregistered unit should be registered now, since the current regulation already requires it.

Is Safety Code 30 law in Manitoba?

Not today. Safety Code 30 is Health Canada guidance, and the X-Ray Safety Regulation does not adopt it. CCMB's wording varies: its owner overview says the codes provide guidance for its operations, while its web page says they set out requirements. The new Act lets regulations adopt a code by reference (section 55(3)). Whether Manitoba will adopt Safety Code 30 is not known. CDHM's interpretation guidelines on dental radiography and on CBCT (both March 2024), which CDHM says hygienists are responsible for complying with, set requirements for hygienists as X-ray equipment operators. Some of those requirements, such as those on thyroid shields and lead aprons, cite Safety Code 30. The MDA's CBCT practice directions list Safety Code 30 only in their references. None of the CDHM or MDA documents we read requires compliance with the code as a whole. We could not read every MDA document, including an MDA E-Alert on Safety Code 30 (January 2024) that CDHM cites.

Sources

Related reading

This page summarizes public sources as of November 1, 2026. It is general information, not legal advice. Confirm your clinic's obligations with the authority named above.