Dental X-Ray Regulations in British Columbia: WorkSafeBC and BCCOHP

The short answer

In British Columbia, WorkSafeBC regulates dental X-ray equipment as a workplace safety matter. Sections 7.23 to 7.25 of the Occupational Health and Safety Regulation (B.C. Reg. 296/97) require three things: equipment installed, operated and maintained to Health Canada Safety Code 30, employer radiation surveys, and survey records kept for at least 10 years. BCCOHP licenses oral health professionals and expects Safety Code 30 compliance.

Who regulates what

X-ray equipment: WorkSafeBC. The Workers' Compensation Board of B.C. (WorkSafeBC) administers the Occupational Health and Safety Regulation under the Workers Compensation Act. BCCOHP names WorkSafeBC as the "applicable regulatory authority" for radiation protection in BC.

Oral health professionals: BCCOHP. The British Columbia College of Oral Health Professionals (BCCOHP) licenses dental hygienists, dental technicians, dental therapists, dentists, denturists and licensed dental assistants. It was established in 2022 by amalgamating BC's legacy oral health colleges. It derives its authority from the Health Professions and Occupations Act, which came fully into force on April 1, 2026.

The national code: Health Canada. Health Canada publishes Safety Code 30 as radiation protection guidance. Where a regulation references a Safety Code, as section 7.23 of BC's Regulation does, Health Canada says that authority determines how compliance is verified.

Associations. The British Columbia Dental Association (BCDA) and the BC Dental Hygienists' Association (BCDHA) are member associations, not regulators.

What the law requires

These duties sit in Part 7, Division 3. Sections 7.22, 7.24 and 7.25 name the employer. "G" numbers are WorkSafeBC's Part 7 guidelines.

Equipment standard (s.7.23(a)(v)). Equipment must be installed, operated and maintained in accordance with Safety Code 30 (1999), "as amended from time to time." Guideline G7.23 says the most recent edition of a listed code applies. For Safety Code 30, that is the 2022 edition.

Radiation surveys (s.7.24). The employer must conduct a survey at the times the Safety Code requires, after equipment is damaged or modified, or when there is an indication of unusually high worker exposure. G7.24 treats moving a unit to a new permanent location as a modification and makes only a survey's worker-protection parts mandatory.

Survey interval (G7.24(a)). The employer sets each unit's interval, weighing factors such as the manufacturer's specifications, shielding, use and age. WorkSafeBC expects it will not normally exceed three years, or five for a typical bitewing unit operating only at or below 70 kVp. BCCOHP restated this guideline, which is not Regulation text, in its December 19, 2025 notice.

Records (s.7.25). Radiation survey records must be kept for at least 10 years and made available to WorkSafeBC and to workers. Exposure monitoring and dosimetry records are kept for employment plus 10 years.

Dosimetry (ss.7.17 and 7.22). Unless WorkSafeBC grants an exemption, the employer must make sure a worker who exceeds or may exceed 1 mSv per year has, and properly uses, a personal dosimeter acceptable to WorkSafeBC.

What these rules do not spell out. Division 3 and the Part 7 guidelines do not themselves list a daily or monthly quality control test for dental units. Safety Code 30 (2022) sets out a schedule of quality control tests, but the guidelines do not say WorkSafeBC enforces it. Ask WorkSafeBC how it applies.

Key dates

  • May 31, 2025: The BCDA X-Ray Inspection Program ended as a member service. BCCOHP said this does not affect its requirements.
  • June 30, 2025: BCCOHP removed from its website the legacy "Dental Radiography" document it lists for licensed dental assistants, dentists and dental therapists.
  • December 19, 2025: BCCOHP published "Expectations for X-ray equipment inspections."
  • April 1, 2026: The Health Professions and Occupations Act came fully into force, and BCCOHP's revised Standards for the Oral Health Team came into force.

A practical checklist for your clinic

  1. Update office procedures that still refer to the BCDA X-Ray Inspection Program. Good practice; the program ended May 31, 2025.
  2. Record each unit's last survey date and interval: normally three years at most, or five for a typical bitewing unit operating only at or below 70 kVp. Guideline G7.24(a).
  3. Book a survey after damage, modification or a move to a new permanent location, or after signs of unusually high staff exposure. Section 7.24(b) and (c); guideline G7.24.
  4. File each specialist's survey report and check it covers the worker-protection items; the employer stays responsible. Guidelines G7.24 and G7.24(a).
  5. Keep survey reports for at least 10 years, available to WorkSafeBC and your staff. Section 7.25.
  6. If any worker exceeds or may exceed 1 mSv per year, provide dosimeters and keep those records for employment plus 10 years. Sections 7.22 and 7.25.

About Phantom XY and Phantom Cloud

Phantom XY and Phantom Cloud help a clinic run its own routine digital QA tests and keep its own records. Together they run an automated digital QA test of about two minutes on intraoral and panoramic X-ray devices, score each result with a Q-Score, and keep the results in one place. Phantom Cloud sends missed-test and failure alerts and includes a Digital X-Ray Logbook. Phantom XY and Phantom Cloud do not perform a radiation protection survey. A test with Phantom XY does not count as a survey and does not change when one is due. The employer still arranges surveys and keeps survey records.

Common questions

Does British Columbia require daily dental X-ray QA?

Not in the regulation's own text. Sections 7.23 to 7.25 of the Occupational Health and Safety Regulation cover equipment installed, operated and maintained to Safety Code 30, employer radiation surveys and 10-year survey records. WorkSafeBC's guidelines say the latest Safety Code 30 edition applies, but they do not say WorkSafeBC enforces that code's daily or monthly tests. Ask WorkSafeBC how they apply to your clinic. Separately, BCCOHP says oral health professionals must meet profession-specific practice standards. Its legacy radiography standard for dental hygienists, which BCCOHP says remains applicable, says a hygienist responsible for radiographic equipment adheres to the parameters of Safety Code 30. That standard does not itself list a daily or monthly test. Ask BCCOHP how it applies.

Who inspects dental X-ray equipment in BC now that the BCDA program has ended?

In BC, the employer is responsible for radiation surveys. The BC Dental Association ended its X-Ray Inspection Program as a member service on May 31, 2025, and BCCOHP says that does not affect its requirements. Under WorkSafeBC guideline G7.24(a), the employer may hire a specialist but remains responsible. WorkSafeBC prevention officers may review survey reports.

How often does a BC dental X-ray unit need a radiation survey?

The employer sets each unit's interval. WorkSafeBC guideline G7.24(a) expects it will not normally exceed three years, or five for a typical bitewing unit operating only at or below 70 kVp. Section 7.24 also requires a survey after damage or modification (G7.24 includes a move to a new permanent location) or an indication of unusually high worker exposure.

How long must a BC dental clinic keep X-ray survey records?

Section 7.25 of the Occupational Health and Safety Regulation requires radiation survey records to be kept for at least 10 years and made available to WorkSafeBC and to workers. Exposure monitoring and dosimetry records are kept for the worker's employment plus 10 years. Under guideline G7.25, if another agency requires a longer period, the longer period prevails.

Does the BC radiation regulation cover the dose a patient receives?

The WorkSafeBC rules focus on workers. Section 7.18(2) says Division 3 does not apply to dental radiation received by a patient, except as WorkSafeBC specifies. Guideline G7.24 makes only the worker-protection parts of a survey mandatory. Separately, BCCOHP expects compliance with Safety Code 30 and points oral health professionals to its standards on legal requirements and safe, hygienic practice environments. Its legacy radiography standard for dental hygienists, which BCCOHP says remains applicable, says a hygienist responsible for radiographic equipment adheres to the parameters of Safety Code 30.

Sources

Related reading

This page summarizes public sources as of November 1, 2026. It is general information, not legal advice. Confirm your clinic's obligations with the authority named above.

Phantom XY and Phantom Cloud help dental teams run a two-minute digital QA test on intraoral and panoramic devices and keep their results in one place. Book a demo

MVD X-Ray Support